medSim 300B Instrument Specifications
ECG Section
The ECG configuration is 12-lead with independent outputs for each signal lead
referenced to RL, and provides:
•
A "High Level ECG" output.
•
Selectable parameters including rate, amplitude, ST segment deviation, axis
deviation, neonatal waveforms, performance waveforms, R-wave detection, PVC
focus, PVC timing, PVCs per minute, arrhythmias (32), pacemaker amplitude,
and pacemaker width.
•
Input for interactive defibrillator training and simulation.
Normal Sinus Rhythm ECG Waveform Selections
Rate:
30, 60, 80, 120, 160, 200, 240, and 300 BPM
Accuracy:
1
%
Amplitude: Lead
II
0.05 mV to 0.5 mV in 0.05 mV steps
0.50 mV to 5.5 mV in 0.25 mV steps
The high level output is approximately 0.2 V per mV of
the low level output. Selection is either manual or
automatic at stepped intervals of 6 seconds
Accuracy: 5 % on Lead II into a 100 Hz low pass filter
Impedance to the RL Lead:
Limb leads are selectable at 500 e, 1000 e, 1500 e and
2000 e. The V-leads are 1000 e
Accuracy:
5%
ST Segment Deviation
(Elevation and Depression): Lead II at 1 mV
+0.8 mV to -0.8 mV in 0.1 mV steps, also +0.05 mV and
-0.05 mV
Both manual and auto-stepping modes. Auto-step time
interval is 12 seconds
Operates at ECG rates from 30 to 160 BPM
Axis Deviation:
Horizontal, vertical and intermediate heart positions