Intrauterine pressure (iup) simulation – Fluke Biomedical medSim 300B User Manual
Page 126

medSim 300B
Operators Manual
4-28
Intrauterine Pressure (IUP) Simulation
The intrauterine pressure (IUP) waveform is output at "BP4/AUX".
The following drawing illustrates the medSim 300B simulated intrauterine pressure
waveform.
0
Seconds
90
10
90
mmHg
eic198.eps
Figure 4-23. medSim 300B Simulated Intrauterine Pressure Waveform
At the blood pressure "SENSITIVITY" switch located on the front panel, adjust the
medSim 300B to match the input sensitivity of the fetal monitor—either 5 µV/V/mmHg
or 40 µV/V/mmHg.
The frequency of the uterine contractions generated by the medSim 300B can be varied.
To vary the rate at which contractions occur:
1. Select
pages of display:
ECG
BASE
PERF
ARRH
>
PVC
eic199.eps
ECG
<
PACE
INS
FETAL
eic200.eps
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)