Power, Power-on conditions, Panel descriptions, controls and connectors – Fluke Biomedical medSim 300B User Manual
Page 32

medSim 300B
Operators Manual
3-4
Parameter Specification
Artifact All
Off
Power-off time
Off
Key beep
On
Baud rate
2400
Cardiac output baseline
37 °C
Note
These power-on conditions can be changed by utilizing the "save" function
in the UTILITY MENU.
Panel Descriptions, Controls and Connectors
Top Panel
LCD Display and Keyboard
The LCD display-2 lines by 24 characters—displays messages, and provides information
necessary for setting output levels and selecting waveform types.
The keyboard has the following eight pressure-sensitive keys: four function keys (F1, F2,
F3, and F4), three arrow keys (left, up and right), and the "ON/OFF" key.
ECG Leads
Located above the LCD display are ten ECG lead-binding posts that are compatible with
disposable snaps, 3.2 mm or 4 mm electrodes, and banana plugs. All leads have
independently generated signals.
RS232
For remote control of the patient simulator, the RS232 serial port interface is used to
connect the medSim Controller Option to a computer or terminal. For information
regarding the operation of the RS-232 interface, see the RS-232 SERIAL PORT
OPERATION section later in this chapter.
High Level ECG
The "HIGH LEVEL ECG" port is a 1/4" standard phone jack with a Lead II waveform at
0.2 V/mV of the ECG Lead II signal. The "HIGH LEVEL ECG" output is designed for
use with central station monitors or recorders requiring an analog high-level input.
Battery Test
Tests 8.4-volt and 9-volt batteries. The LED illuminates when the battery has more than a
6.2-volt charge.
Lead Test
Tests continuity of the ECG leads. The LED illuminates when the resistance is less than
1 ke.