Fluke Biomedical medSim 300B User Manual
Page 78

medSim 300B
Operators Manual
3-50
Function Select - Sequences -Step-sequences - User Sequences Menu
After programming user step-sequences, from the USER SEQUENCES MENU choose
the user step-sequence to run.
To select the USER SEQUENCES MENU:
1. Press
display:
USER SEQUENCES
SEQ1
SEQ2
SEQ3
SEQ4
eic120.eps
2. Press the corresponding function key to select the desired sequence.
Note
If a selected sequence has not yet been programmed, an error beep sounds.
To program user step-sequences, follow the instructions that appear in the
Make Sequences Menu section located earlier in this chapter.
Upon pressing a function key related to a previously programmed user step-sequence, the
display, as represented below, provides the following information:
S# ##:##.#
????
##.#
strt
stop
next
rset
eic121.eps
From the top line of the display:
S#
Is the sequence name: "S1", "S2", "S3", or "S4"
##:##.#
Is the sequence run time
????
Is the segment name (abbreviated)
##.#
Is the segment run time
Menu operations for user step-sequences are the same as for preset (or fixed)
step-sequences.
The user step-sequence selections and descriptions follow:
Selections Descriptions
strt
Starts the sequence running
stop
Stops the sequence
next
Steps to the next segment if running
rset
Reset stops the sequence, returns to the first segment, and clears the
time. Reset does not restart the sequence