Fluke Biomedical medSim 300B User Manual

Page 58

background image

medSim 300B

Operators Manual

3-30

2. Press the corresponding function key for the desired BP 1 dynamic waveform.

Waveform selections and their descriptions follow.

Selections Descriptions
atm

Atmosphere - 0 mmHg.

art

Arterial at 120/80 mmHg.

lv

Left ventricle at 120/0 mmHg.

cvp

Central venous pressure at 15/10 mmHg.

rv

Right ventricle at 25/0 mmHg.

pa

Pulmonary artery at 25/10 mmHg.

paw

Pulmonary artery wedge at 10/2 mmHg.

To select BP 1 static pressures, press -STAT from the BP 1 MENU for the

following display:

bp 1 static:

## mmHg

down

up

exc

auto

eic073.eps005A

Follow the instructions for the step menu functions presented earlier in this chapter.
BP 2 dynamic waveforms and static pressures are identical to BP 1. To select BP 2, press

-2 from the BP CHANNEL SELECT MENU to access the following menu:

BP 2

WAVE

STAT

eic074.eps

To select a BP 2 dynamic waveform:

1. Press -WAVE to access the two menu pages as shown:

bp 2 waveforms

atm

art

lv

cvp

>

eic075.eps