Pulmonary valve simulation – Fluke Biomedical medSim 300B User Manual

Page 108

background image

medSim 300B

Operators Manual

4-10

The single lumen catheter pullback (aortic valve) simulation begins as follows on

the medSim 300B:

bp2:art

ins

bp1:art

eic164.eps

The top line of the display gives the following information:

bp2:art

BP channel 2 arterial blood pressure—120/80

bpl:art

BP channel 1 arterial blood pressure—120/80

2. To simulate insertion of the catheter back into the left ventricle select -INS.

The medSim 300B returns to the previous display as shown below:

bp2:art

plbk

down

up

bp1:lv

###/0

eic165.eps

Note

In the above display "###/0" represents the previously selected left

ventricular pressure value.

3. To exit the medSim 300B aortic valve simulation, press the key until

you view the desired menu.

Pulmonary Valve Simulation

For pulmonary valve simulation, using medSim 300B channels BP3 and BP4/AUX,

connect the medSim 300B to the cardiac catheterization system under test.