Pulmonary valve simulation – Fluke Biomedical medSim 300B User Manual
Page 108

medSim 300B
Operators Manual
4-10
The single lumen catheter pullback (aortic valve) simulation begins as follows on
the medSim 300B:
bp2:art
ins
bp1:art
eic164.eps
The top line of the display gives the following information:
bp2:art
BP channel 2 arterial blood pressure—120/80
bpl:art
BP channel 1 arterial blood pressure—120/80
2. To simulate insertion of the catheter back into the left ventricle select
The medSim 300B returns to the previous display as shown below:
bp2:art
plbk
down
up
bp1:lv
###/0
eic165.eps
Note
In the above display "###/0" represents the previously selected left
ventricular pressure value.
3. To exit the medSim 300B aortic valve simulation, press the
you view the desired menu.
Pulmonary Valve Simulation
For pulmonary valve simulation, using medSim 300B channels BP3 and BP4/AUX,
connect the medSim 300B to the cardiac catheterization system under test.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)