Technical Data
6-3
Positioner SIPART PS2/SIPART PS2 PA
C79000-G7476-C150–01
Electrical data: basic instrument without PROFIBUS PA
6DR4000–1N/–2N and
6DR4001–1N
6DR4000–1E/–2E and
6DR4001–1E
Degree of protection acc. to
EN 50014 and EN 50020
–
II2G EEx ib II C T4/T5/T6
Place of installation
–
zone 1 + 2
Ambient temperature
– 30 to + 80
°
C
T4 – 30 to + 80
°
C
T5 – 30 to + 65
°
C
T6 – 30 to + 50
°
C
2-wire circuit
Fig. 3-8
Fig. 3-16
Nominal signal range
4 to 20 mA
Current for maintenance of
power supply
w
3.6 mA
Required load voltage
U
B
w
10 V without HART module
U
B
w
11.5 V with HART module
Static destruction limit
"
40 mA
–
Internal capacitance
C
i
–
19 nF
Internal inductance
L
i
–
0.22 mH
For connection to intrinsically
safe current sources
with
U
i
–
v
30 V
J
i
–
v
100 mA
P
i
–
v
1 W
3/4-wire circuit
Fig. 3-9
Power supply U
H
Range
+ 18 to 35 V
+ 18 to 30 V
Power consumption J
H
Static destruction limit
"
35 V
–
Internal capacitance
C
i
–
19 nF
Internal inductance
L
i
–
0.22 mH
For connection to intrinsically
safe current sources
with
U
i
–
v
30 V
J
i
–
v
100 mA
P
i
–
v
1 W
U
H
– 9.2 V
J
H
=
1.5 k
Ω
[mA]