Center-of-zero (coz) indicator, Reset all, Are you sure – Rice Lake MSI-9850 CellScale RF Digital Indicator - Operator Manual User Manual
Page 94: Msi-9850

Page 94 MSI-9850 RF Remote Indicator • User Guide
MSI-9850
HANDHELD RF REMOTE INDICATOR for
C
ELL
S
CALE
®
CENTER-OF-ZERO (COZ) INDICATOR
The COZ indicator turns on when the scale is within 1/4 d of the last zero setting. Some legal-for-trade jurisdic-
tions require its use.
To Enable/Disable the COZ Indicator
Before starting the following procedure, you must be in the Top Menu Level of Calibration.
1) Press
SETUP
, then
CHANNEL
.
2) Select “Calibration Settings”
[3]
.
3) Toggle the Center of Zero indicator on or off with the
[4]
key.
4) Press
DISPLAY
to return to the Scale Display. Press
ESC
to return to the “CHANNEL / CALIBRATION”
menu.
RESET ALL
This procedure returns all registers of the 9850 to defaults. DO NOT initiate this function unless you are prepared
and qualified to perform a complete setup of the 9850 system. This operation does not clear calibrations or affect
any connected CellScales.
To Completely Reset the 9850
To return to the
startup screen
To return to the
weight display
To Clear All Setups
back to defaults
To cancel the
Clear to defaults
To Reset the RF Modem
back to defaults
Len 0 F4 Menu
Max 14 ENT Save
Pos 1
ENTER PASSWORD
ARE YOU
SURE
POWER
Any Key
If a
Password
is Set
No
Password
ENTER to
Clear All
SEND to
Reset RF
ESC
!
SETUP
ZERO
CLR
0
ESC
!
SETUP
DISPLAY
EXIT
IN
F3
SEND/PRINT
ENTER
DISPLAY
EXIT
1) Press
POWER
followed quickly by
ESC
(Setup). The MSI-9850
startup screen will appear.
2) Press the
CLR
(
ZERO
) key. If a password is set, you must enter it
in the “ENTER PASSWORD” screen.
3) The “ARE YOU SURE” screen appears. Press
ENTER
to reset the
9850. All 9850 parameters are returned to defaults.
4) After the Clearing function is complete, the startup screen will appear
again. Press
DISPLAY
to return to the weight display mode. Only
channel 1 of the current Network will display until the 9850 is setup
for more.
Go to Index