Msi-9850, Set points, Driving 5v coil relays – Rice Lake MSI-9850 CellScale RF Digital Indicator - Operator Manual User Manual
Page 16: Driving externally powered coil relays
![background image](/manuals/639371/16/background.png)
Page 16 MSI-9850 RF Remote Indicator • User Guide
MSI-9850
HANDHELD RF REMOTE INDICATOR for
C
ELL
S
CALE
®
9850 P3
Set Points Port
Pin 1- 5V Out
Pin 2 - SP2
Pin 3 - SP3
Pin 4 - SP5
or GND
Pin 5 - SP1
Set Points
BRN
WHT
BLUE
BLK
GREY
5Vdc
Snubber Diodes
not required
Driving 5V Coil Relays
Limit 150mA ea
max coil current
(or 500mA Total)
+
+
+
+
Pin 5 - SP4
PINK
+
Use of SP5
requires Jumper
JP6 positioned
1-2. JP6 is
defaulted to GND
and will turn on
any relay
connected to the
black wire when
the 9850 power
is turned on.
Insulate all unused wires.
9850 P3
Set Points Port
Pin 1- 5V Out
Pin 2 - SP2
Pin 3 - SP3
Pin 4 - GND
Pin 5 - SP1
Set Points
BRN (NC)
WHT
BLUE
BLK
GREY
5Vdc-25Vdc
Snubber Diodes
not required
Driving Externally Powered Coil Relays
Limit 150mA ea
max coil current
Pin 5 - SP4
PINK
Insulate Brn
Wire
Insulate all unused wires.
JP6 in default GND
Position (1-4)
Relay
Power
Source
+
–
+
+
+
+
Brown Provides a +5V source for driving relays or auxiliary devices. Fused with a self resetting Polyswitch
device. Maximum output current is 500mA total. This output is diode protected and not highly regu-
lated as the regulated voltage passes through a Schottky diode. Since V-forward of the Schottky diode
varies with current drain, the voltage will vary from a low of ~4.9V at 500mA to a high of 5.2V lightly
loaded.
White Open Drain Relay Driver, Set Point 2
Blue
Open Drain Relay Driver, Set Point 3
Black Defaults to circuit ground. Power return for the 5V supply. This line can be used to drive a master relay
that turns on when the 9850 turns on. Or by moving JP6 to poisition 1-2, provides an Open Drain Relay
Driver, Set Point 5
Grey Open Drain Relay Driver, Set Point 1
Pink
Open Drain Relay Driver, Set Point 4
With 5V relays, no other
power supply is required for
driving relay coils. Other
5V devices such as lamps or
alarms can also be powered
by these lines as long as the
maximum current rating is
observed.
Driving higher volt-
age devices (up to 25V)
requires the use of an exter-
nal supply. JP6 should be in
its default position 1-4 to
provide a ground reference
to the relay power source.
Therefore only 4 outputs
(SP1-4) are available.
When the 9850 is DC pow-
ered, SP5 can be used since
the ground reference is
provided through the power
input.
9850 P3
Set Points Port
Pin 1- 5V Out
Pin 2 - SP2
Pin 3 - SP3
Pin 4 - SP5
Pin 5 - SP1
Set Points
BRN (NC)
WHT
BLUE
BLK
GREY
7Vdc-25Vdc
Snubber Diodes
not required
Driving Externally Powered Coil Relays
Limit 150mA ea
max coil current
Pin 5 - SP4
PINK
Insulate Brn
Wire
Insulate all unused wires.
JP6 in Position (1-2)
9850 &
Relay
Power
Source
+
–
+
+
+
+
+
To 9850 DC Power
Input. Negative
Power shared.
Go to Index