Page 50 MSI-9850 RF Remote Indicator • User Guide

MSI-9850

HANDHELD RF REMOTE INDICATOR for

C

ELL

S

CALE

®

To Set the Total Mode

The following example assumes the F1 function key is in its default mode of “TOTAL”.

To Turn On or Off Auto Total

1) Press

SETUP

.

2) Press the

TOTAL

switch

[F1]

or select

[6]

Total.

3) The current Total Mode is shown on selection 1.
4) Press the

[1]

key to bring up the “Total Mode Selection” menu.

5) Select the desired Total Mode with the numeric keys (

1

-

7

). You are returned to the “Total Status” with the

new mode indicated on selection 1.

6) Motion Setting – If totals are allowed before the scale settles, disable Motion by pressing

[2]

. Disabling

Motion should only be used in situations where the weight may never settle. Accuracy will suffer.

Some AutoTotal modes require no motion to function (Auto Load, Auto Norm) and will ignore the setting here. ON means totals will not function until the
weight has settled. OFF indicates that motion conditions are ignored.

7) Press the

EXIT

key to return to normal scale operation or the

ESC

key to continue in the Setup Menus.

Turn on or off Auto Totaling by pressing the

TOTAL

key

[F1]

. The display will read “TOTAL ON” along

with the current total mode, or “TOTAL OFF”.

To Set the Total Threshold

VIEW TOTAL

Displays the current Total value of the selected ID and allows the total(s) to be cleared.

The following procedures assume the F5 function key is programmed to “VIEW TOTAL” instead of “UNITS”. In any View Total screen, you can change channels
by pressing the CHANNEL key. You can change to the next ID by pressing the ID key.

To Display the Current ID#’s Totaled Weight and Statistics

1) Press

SETUP

.

2) Press the

TOTAL

switch

[F1]

or select

[6]

Total.

3) Select “High Threshold” by pressing

[3]

.

4) Enter the desired high threshold with the numeric keys. Threshold is expressed in per cent of capacity.
5) Set the new high threshold by pressing

ENTER

.

If the high threshold you entered was lower than the previous low threshold, the CellScale automatically adjusts the low threshold to be 1% less than
the high threshold.

6) To change the low threshold, press the

[4]

key.

7) Enter the desired low threshold with the numeric keys.
8) Set the new low threshold by pressing

ENTER

.

If the low threshold you entered was higher than the previous high threshold, the CellScale automatically adjusts the high threshold to be 1% more than
the low threshold.

9) Press the

EXIT

key to return to normal scale operation.

1) Press

[F5]

. The current IDs total appears on line two, the number of

totaled samples on line three, and the ID name (if programmed) on line

1.

2) To return to the standard weight display, press

DISPLAY

or

ESC

.

ID NAME ID# CH#

1000.5lb ∑

6 #

1 Grand 3 Clr Last
2 Stats CLR Delete

1) Press

VIEW

. The current total is displayed.

2) Select “Clr Last” by pressing

[3]

. The “ARE YOU SURE” screen gives an opportunity to change your

mind.

3) To complete the last totaled weight deletion, press the

ENTER

key.

To Erase the last Totaled Weight

If the last totaled (automatic or manual) weighment was a mistake, it can be erased with the following procedure. This erases only the last totaled weighment.