You can also change the Backlight settings using the UP/DOWN cursor keys to select the parameter, and the LEFT/RIGHT cursor keys to change the value.
Setting the LCD Contrast
LCD contrast is effected by Temperature, humidity, and viewing angle. The 9850’s LCD is temperature compen-
sated and usually does not require adjustment due to temperature variations. Only extremely cold (<0°C) or hot
conditions (>50°C) might require manual contrast adjustment. Viewing angle, however, is a matter of taste. The
9850 sets its fine contrast adjustment at a setting of 10. Setting the contrast lower than 10 will cause the display
to darken and turn more purple. Setting it higher than 10 will lighten the display. Set the contrast while holding
the 9850 at your usual operational angle.
Method 1 Cursor Keys
1) Press
SETUP
,
DISPLAY
.
2) Press the down cursor key
[F5]
5 times to highlight “Contrast”.
2) Press the right cursor key
[F6]
to lighten the display or use the left cursor key
[F2]
to darken the display.
3) Press
ESC
to backup to the SETUP SELECT MENU. Press
ESC
again to return to the weight display.
Method 2 Numeric Entry
1) Press
SETUP
,
DISPLAY
.
2) Select “Contrast” by pressing [5].
3) Using the numeric keypad, enter a contrast value. Although the number entry screen takes numbers from
0-63, the usable contrast values are about 5 to 20.
4) You will not see the contrast change until you press
ENTER
. At this point you can use the Left/Right Cursors
to fine tune the contrast.
5) Press
ESC
to backup to the SETUP SELECT MENU. Press
ESC
again to return to the weight display.
STANDARD & CUSTOM DISPLAY SETUP
The 9850 offers the user a great deal of flexibility in how weight and other information is displayed. There are
five single channel preset displays and five multiple channel preset displays. These presets cover the majority
of user applications. However, the 9850 also allows complete customizing of how scale data is presented on the
display.
There are two basic weight display modes, Single Channel, and Multiple Channel. The single channel display is
the primary weight display, and is used as the display mode when a channel is highlighted and switched out of
the multiple channel display. If your CellScale is equipped with multiple channels, the display can be set up to
show up to eight channels at once.
Preview Screens
When setting up Weight Displays, the 9850 shows a preview of how the screen will be formatted. Abbreviations
for preview screens indicate what kind of data will be presented in each screen location. Dashes are used to indi-
cate the maximum screen space the data will occupy.
Preview Screen Nomenclature
Weight ................ Current Weight in Net or Gross with Units, Motion, and Center of Zero (COZ) indicators.
Gross-Wt ........... Gross Weight only
Tare-Wt .............. Tare Weight only
CHN .................. Channel number in the form X-X. e.g. 1-1. The Channel number is determined by the CellScale
Scan List.
ADR ................. Network Address in the form of XXA. e.g. 17A
Ch-Name ............ Channel Name. The Channel Name is programmed in the CellScale.
DateTime ............ Shows the Date and Time. When in the form “DateTimeEnter-Number” it may be overwritten
when entering a number with the keypad.
Enter-Number ..... Field accepts keypad entry for assignment to various functions.
1-ID-String ......... Show ID string 1.
2-ID-String ......... Show ID string 2.