Channel / calibrate settings, Msi-9850 – Rice Lake MSI-9850 CellScale RF Digital Indicator - Operator Manual User Manual
Page 106
![background image](/manuals/639371/106/background.png)
Page 106 MSI-9850 RF Remote Indicator • User Guide
MSI-9850
HANDHELD RF REMOTE INDICATOR for
C
ELL
S
CALE
®
CHANNEL FILTER
1 Off
2 Low
3 High
1 Channel INACTIVE
1 Channel is ACTIVE
CALIBRATE -CHN-
SETTINGS
1 Motion Band 0d
2 Standard INDUSTRY
3 Center of Zero OFF
4 AutoZero (AZM) OFF
5 AZM Band 1d
Starts the Calibration
Procedure
3 Center of Zero OFF
3 Center of Zero ON
SET MOTION BAND (d)
ESC exits no change
ENTER saves value
0-9 replaces value
10
min = 0 max = 255
^ incs, v decs digit
SCALE CAL STANDARD
1 Industrial (unreg)
2 NIST NTEP
3 OIML R76
4 Industrial Metric
4 AutoZero (AZM) OFF
4 AutoZero (AZM) ON
AUTOZERO BAND (d)
ESC exits no change
ENTER saves value
0-9 replaces value
10
min = 0 max = 255
^ incs, v decs digit
5 Zero RUSure Msg OFF
5 Zero RUSure Msg ON
CHANNEL / CALIBRATION
--Channel Name--
--Chn-- --Index--
1 Channel is ACTIVE
2 Filter LOW
3 Calibrate Settings
4 Start Calibration
5 Zero RUSure Msg ON
Changes to the
next Channel
available
Shortcut
CHANNEL / CALIBRATION
--Channel Name--
--Chn-- --Index--
CALIBRATE 1-1
SETTINGS
CHANNEL FILTER
SET MOTION BAND (d)
SCALE CAL STANDARD
AUTOZERO BAND (d)
ESC
!
SETUP
2
DEF
1
ABC
3
GHI
5
MNO
4
JKL
7
STU
F2
CHANNEL
1
ABC
2
DEF
3
GHI
4
JKL
1
ABC
F2
CHANNEL
5
MNO
CHANNEL / CALIBRATE SETTINGS