Maintenance and Troubleshooting 10-17
Problem
Monitor establishes a connection, but does not respond
to any message.
Possible
Causes
Cabling may be loose.
Lightning protection module may be damaged.
Modem in monitor may be faulty.
Possible
Solutions
Listen for noise at the site using a field phone. If noise is
present, inspect the wirings and replace wiring if
necessary.
Replace the lightning protection module.
Contact telephone company.
Collect the data from the monitor on site using the serial
cable, and replace monitor if defective.
Contact your regional ADS representative.
Problem
Time stamp on the collected data is incorrect.
Possible
Causes
PC clock may read incorrect time.
Monitor clock may be faulty.
Possible
Solutions
Verify the time on the PC clock and correct if necessary.
Reactivate the monitor to enable the clock.
Collect the data from the monitor and replace monitor if
defective.