Analog signal for HDA 5500

HDA status signal 5 table

The current or voltage of the output signal (5) is dependent on the status of the
CS1000 as shown in the table below:

Current I

Status

Voltage U

I = 5.0 mA

CS is functioning correctly

U = 2.5 V

I = 6.0 mA

Device error / CS not ready

U = 3.0 V

I = 7.0 mA

Flow insufficient (Flow 2 low)

U = 3.5 V

I = 8.0 mA

ISO <9.<8.<7

U = 4.0 V

I = 9.0 mA

No measured value

(flow not defined)

U = 4.5 V

If the status signal is ≥ 6.0 mA or ≥ 3.0 V, signals 1 to 4 are output with 20 mA or 10