Exception errors – HYDAC CS 1000 User Manual
Page 80

ContaminationSensor CS1000
Status messages
HYDAC Filtertechnik GmbH
en
Page 80 / 100
BeWa CS1000 3247149n en 2009-02-04.doc
2009-02-04
LED Blink code /
Display /
Analog output /
SwitchOut
Status
To do
Error
No.
Red
The sensor is above
its measurement
range ISO 25/24/23 Clean/purify oil
3
Exception errors
LED
Blink code /
Display /
Analog output /
SwitchOut
CS 1000 Status
To do
Error
No.
Out
0 mA / 0 V
Open
CS
no display
no function
Check supply voltage
for CS
Contact HYDAC
-
Red
4.1 mA / 2.05 V* or
19.9 mA / 9.95 V
Open
"2 low" on "Drive"
If the CS is being
operated with 24 V,
then connect it to a
12 V voltage supply
or contact HYDAC
-
*
Is not valid for HDA 5500 output signal (see Table 0)