Using switching output

HYDAC Filtertechnik GmbH

en

Page 45 / 100

BeWa CS1000 3247149n en 2009-02-04.doc

2009-02-04

Using switching output

A description is provided below as to how the switch output behaves in the various
modes and thus how it can be operated by the user.

For a further description of the measurement modes, see also Chapter "CS
Measurement modes".

Mode M1: Continuous measurement

Purpose: Measurement

only

Function:

measurement of solid contamination, without without any switching
functions

Mode M2: Continuous measurement and switching

Purpose:

Continuous measurement and controlling of signal lamps etc.

Function:

Continuous measurement of solid contamination, continuous
monitoring of programmed limit values; the switching output is
enabled and switches on the monitoring display or alarm on site

Mode M3: Filter to cleanliness class and stop

Purpose:

Clean up hydraulic reservoir

Function:

Control of a filter unit, continuous measurement of solid
contamination; if pre-programmed cleanliness level is reached 5
times in sequence, the pump is stopped.

Mode M4: Filter to continuously monitor cleanliness class

Purpose:

Establish continuous monitoring of cleanliness class between
min/max limit values

Function:

If min/max limit values are pre-programmed, the CS switches the
filter unit on/off to keep cleanliness within the limit value range

"SINGLE" measurement mode

Purpose:

Perform a single measurement and "stop" the result.

Function:

Single measurement of solid contamination without switching
functions