CS measurement modes
After powering up, the CS1000 automatically runs in the measuring mode that has
been set.
Mode M1: Continuous measurement
Application: Stand-alone
sensor
Data output:
Display & RS485 & analog output
Purpose: Measurement
only
Function:
Continuous measurement of cleanliness class, without
switching functions
Mode M2: Continuous measurement and switching
Application:
Stand-alone sensor with alarm standby display
Data output:
Display & RS485 & analog output & switching output
Purpose:
Continuous measurement and controlling of signal lamps etc.
Function:
Continuous measurement of solid contamination, continuous
monitoring of programmable limit values; switching output is
activated to switch on the monitor display or an alarm on site
Mode M3: Filter to cleanliness class and stop
Application:
Controlling a filter unit
Data output:
Display & RS485 & analog output & switching output
Purpose:
For cleaning up a hydraulic reservoir
Function:
Controlling a filter unit, continuous measurement of solid
contamination. If pre-programmed cleanliness level is achieved
5 times in sequence, the pump is stopped.
Mode M4: Filter to continuously monitor cleanliness class
Application:
Control of stationary offline filtration unit
Data output:
Display & RS485 & analog output & switching output
Purpose:
Establish continuous monitoring of cleanliness class between
min/max limit values
Function:
Controlling a filter unit, continuous measurement of solid
contamination.
If min/max limit values are pre-programmed, the CS switches
the pump on/off to keep cleanliness within the limit value range
Mode "SINGLE" individual measurement