ContaminationSensor CS1000

Analog Output (ANAOUT)

HYDAC Filtertechnik GmbH

en

Page 57 / 100

BeWa CS1000 3247149n en 2009-02-04.doc

2009-02-04

Current I

ISO contamination class / error

Voltage U

I = 17.90 mA

ISO 23

U = 8.95 V

I = 18.47 mA

ISO 24

U = 9.24 V

I = 19.20 mA

ISO 24.28

U = 9.60 V

19.2 mA < I < 19.8 mA

Not defined

9.60 V < U < 9.90 V

19.8 mA < I < 20 mA

No measured value

9.90 V < U < 10 V

The current (I) or voltage (U) can be calculated for a given ISO contamination

class as follows:

I = 4.8 mA + ISO code x (19.2 mA - 4.8 mA) / 24.28

U = 2.4 V + ISO Code x (9.6 V - 2.4 V) / 24.28

The ISO contamination class can be calculated for a given current I or voltage U

as follows

ISO code = (I - 4.8 mA) x (24.28 / 14.4 mA)

ISO code = (U - 2.4 V) x (24.28 / 7.2 V)