ContaminationSensor CS1000
Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 74 / 100
BeWa CS1000 3247149n en 2009-02-04.doc
2009-02-04
HDA.NAS – Analog Signal NAS to HDA 5500
The HDA.NAS signal consists of 4 measured values (Start / NAS 2 / NAS 5 / NAS
15 / NAS 25 / Status) which are output sequentially. Synchronization with the
downstream control unit is a prerequisite.
The signal output is as follows:
Time
Measurement
category
Signal duration
in s
Current/Voltage
Start signal 0
--
2
20 mA / 10 V
Pause
2
4 mA / 2 V
Signal 1
2-5 µm
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 2
5-15 µm
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 3
15-25 µm
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 4
> 25 µm
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 5
Status
2
Current/voltage for signal
Pause
30
4 mA / 2 V