ContaminationSensor CS1000
Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 59 / 100
BeWa CS1000 3247149n en 2009-02-04.doc
2009-02-04
ISO + T (ISO+T)
The ISO+T signal consists of 4 measured values which are transmitted time-coded
with the following time slices:
4,0
2,0
4,1
2,05
I (mA)
U (V)
time (ms)
4,8
2,4
19,2
High
High
Low
Low
9,6
300
3000
3000
3000
1
2
4
6
8
3
5
7
19,5
9,85
4,3
2,15
19,7
9,75
0,0
19,8
9,9
Ti
me
Signal
Size
Signal duration
per pulse
in ms
Current (I) / Voltage (U)
1
Identifier
> 4 µm(c)
300
High / Low
2
Measured value
> 4 µm(c)
3000
Current/Voltage for measured value
3
Identifier
> 6 µm(c)
300
High / Low / High / Low
4
Measured value
> 6 µm(c)
3000
Current/Voltage for measured value
5
Identifier
> 14 µm(c)
300
High / Low / High / Low / High / Low
6
Measured value
> 14 µm(c)
3000
Current/Voltage for measured value
7
Identifier
Temperature
300
High / Low / High / Low / High / Low / High /
Low / High / Low
8 Measured
value Temperature
3000
Current/Voltage for measured value