Status messages, Status led / display, Error – HYDAC CS 1000 User Manual
Page 79

ContaminationSensor CS1000
Status messages
HYDAC Filtertechnik GmbH
en
Page 79 / 100
BeWa CS1000 3247149n en 2009-02-04.doc
2009-02-04
Status messages
Status LED / display
LED Blink code /
Display /
Analog output /
Switch out
Status
To do
Error
No.
Gree
n
Conductive
CS o.k.
---
-
Red
Current value mA / V*
Conductive
The sensor is below
its measurement
range ISO 9/8/7
---
-
Error
LED Blink code /
Display /
Analog output /
SwitchOut
Status
To do
Error
No.
Red
4.4 mA / 2.2 V*
Open
Flow insufficient
Check that the flow is
between 30 - 300
ml/min.
Increase inlet
pressure or reduce
outlet pressure.
1
Red
19.9 mA / 9.95 V*
Open
Flow definition not
possible
Sensor status is
undefined
Check flow
When the oil
cleanliness is below
the measurement
limit (ISO 9/8/7), it
can take a few
measuring cycles
until measured
values are displayed
3