Fluid temperature (temp) – HYDAC CS 1000 User Manual

Page 77

background image

ContaminationSensor CS1000

Analog Output (ANAOUT)

HYDAC Filtertechnik GmbH

en

Page 77 / 100

BeWa CS1000 3247149n en 2009-02-04.doc

2009-02-04

Fluid temperature (TEMP)

The current range 4.8 - 19.2 mA or voltage range 2.4 - 9.6 V is dependent on the
fluid temperature of from -25°C – +100°C (resolution: 1°C) or -13°F – 212°F
(resolution: 1°F)

Current I

Temperature / error

Voltage U

I < 4.00 mA

Cable break

U < 2.00 V

4.0 mA < I < 4.1 mA

Device error, device not ready

2.00 V < U < 2.05 V

4.1 mA < I < 4.3 mA

Not defined

2.05 V < U < 2.15 V

4.3 mA < I < 4.5 mA

Flow error (flow insufficient)

2.15 V < U < 2.25 V

4.5 mA < I < 4.8 mA

Not defined

2.25 V < U < 2.40 V

I = 4.8 mA

-25 °C / -13 °F

U = 2.40 V

...

...

...

I = 7.68 mA

0 °C / 32 °F

U = 3.84 V

I = 8.26 mA

+5 °C / 41 °F

U = 4.13 V

I = 8.83 mA

+10 °C / 50 °F

U = 4.42 V

I = 9.41 mA

+15 °C / 59 °F

U = 4.70 V

I = 9.98 mA

+20 °C / 68 °F

U = 4.99 V

I = 10.56 mA

+25 °C / 77 °F

U = 5.28 V

I = 11.14 mA

+30 °C / 86 °F

U = 5.57 V

I = 11.71 mA

+35 °C / 95 °F

U = 5.86 V

I = 12.29 mA

+40 °C / 104 °F

U = 6.14 V

I = 12.86 mA

+45 °C / 113 °F

U = 6.43 V

I = 13.44 mA

+50 °C / 122 °F

U = 6.72 V

I = 14.02 mA

+55 °C / 131 °F

U = 7.01 V

I = 14.59 mA

+60 °C / 140 °F

U = 7.30 V

I = 15.17 mA

+65 °C / 149 °F

U = 7.58 V

I = 15.74 mA

+70 °C / 158 °F

U = 7.87 V

I = 16.32 mA

+75 °C / 167 °F

U = 8.16 V

I = 16.90 mA

+80 °C / 176 °F

U = 8.45 V

I = 17.47 mA

+85 °C / 185 °F

U = 8.74 V

I = 18.05 mA

+90 °C / 194 °F

U = 9.02 V

I = 18.62 mA

+95 °C / 203 °F

U = 9.31 V

I = 19.20 mA

+100 °C / 212 °F

U = 9.60 V

19.2 mA < I < 19.8 mA

Not defined

9.60 V < U < 9.90 V

19.8 mA < I < 20 mA

No measured value

9.90 V < U < 10 V