ContaminationSensor CS1000

Analog Output (ANAOUT)

HYDAC Filtertechnik GmbH

en

Page 67 / 100

BeWa CS1000 3247149n en 2009-02-04.doc

2009-02-04

HDA.ISO – Analog signal ISO to HDA 5500

The HDA.ISO signal consists of 4 measured values (ISO 4 / ISO 6 / ISO 14 / ISO 21
/ Status) which are output sequentially. Synchronization with the downstream control
unit is a prerequisite.

The signal output is as follows:

Time

Measurement

category

Signal duration

in s

Current/Voltage

Start signal 0

--

2

20 mA / 10 V

Pause

2

4 mA / 2 V

Signal 1

> 4 µm

2

Current/Voltage for signal

Pause

2

4 mA / 2 V

Signal 2

> 6 µm

2

Current/Voltage for signal

Pause

2

4 mA / 2 V

Signal 3

> 14 µm

2

Current/Voltage for signal

Pause

2

4 mA / 2 V

Signal 4

> 21 µm

2

Current/Voltage for signal

Pause

2

4 mA / 2 V

Signal 5

Status

2

Current/voltage for signal

Pause

30

4 mA / 2 V