OPERATION

6

99

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

6.5.11 System control

Menu C1.4.1 - Function

This menu makes it possible to switch off certain measuring functions depending on a
programmable process state. When this system state occurs (as set in Fct. C1.4.2), the following
options can be activated:

• Inactive: system control is off
• Flow = 0: Flow set to zero

Menu C1.4.2 - Condition

Selecting the process measurement which activates the system control. Density and
temperature can be selected.

Menu C1.4.3 – Max Limit
Menu C1.4.4 – Min Limit

Setting the limit values to activate system control. Current measuring values outside of this
range activate this function.