Operation – KROHNE MFC 400 EN User Manual

Page 80

background image

6

OPERATION

80

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

C3._ Limit Switch X

C3._

Limit Switch X

X stands for one of the connection terminals A, B, C or D

_ stands for Fct. C3.2 (A) / C3.3 (B) / C3.4 (C) / C3.5 (D)

C3._.1

Measurement

Select:

Volume Flow / Mass Flow / Temperature / Density / Velocity / Sensor

Average / Sensor Deviation / Drive Level / Tube Frequency / Strain 1 /

Strain 2 / 2 Phase Signal

C3._.2

Threshold

Switching level, set threshold with hysteresis

xxx.x ±x.xxx (format and unit depend on the measurement, see above)

(1st value = threshold / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C3._.3

Polarity

Set polarity; the flow direction in C1.3.2 must be noted!

Select:

both polarities (plus and minus values are displayed) /

Positive Polarity (display for negative values = 0) /

Negative Polarity (display for positive values = 0) /

Absolute Value (always displays positive, with both negative and positive

values)

C3._.4

Time Constant

Range: 000.1…100 s

C3._.5

Invert Signal

Select:

Off (activated output: switch closed) /

On (activated output: switch open)

C3._.6

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C3._.7

Simulation

Sequence see B1._ Limit Switch X

C3._ Control Input X

C3._

Control Input X

C3._.1

Mode

X stands for connection terminal A or B

_ stands for Fct. C3.2 (A) / C3.3 (B)

Off (control input switched off) /

Hold All Outputs (hold actual values, not display and totalisers) /

Hold Output Y (hold actual values) /

All Outputs To Zero (actual values = 0%, not display and totalisers) /

Output Y To Zero (actual value = 0%) /

All Totalisers (reset all totalisers to "0") /

Totaliser "Z" Reset (set totaliser 1, 2 or 3 to "0") /

Stop All Totalisers /

Stop counter "Z" (stops counter 1, 2 or 3) /

Zero Outp.+Stop Tot. (all outputs 0%, stop all totalisers, not the display) /

External Range Y (control input for external range of current output Y) - also

make this setting on current output Y (no check if current output Y is

available) /

Error Reset (all resettable errors are deleted) /

Calibrate Zero

C3._.2

Invert Signal

Select:

Off (activated output: switch closed) /

On (activated output: switch open)

C3._.3

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C3._.4

Simulation

Sequence see B1._ Control Input X

.book Page 80 Monday, August 26, 2013 9:05 AM