Service, 2 sensor or drive coil fault – KROHNE MFC 400 EN User Manual

Page 117

background image

SERVICE

7

117

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

7.2 Sensor or drive coil fault

The specified values are for guidance only.

Typical values for sensor or drive coils

Typical RTD resistances

• Resistance values outside of the ranges specified above may indicate a circuit error. The

device may be in "start-up" mode or indicate measuring errors.

• All circuits should be isolated with respect to earth (device housing) and each other >20 MΩ.
• A drive/sensor coil short circuit can cause the device to be put into "Startup mode".

Size

Resistance (Ohm)

Driver

Sensor A - Sensor B

black/grey

purple/green - white/yellow

Cryo 230°C /
446°F

400°C / 752°F

Cryo 230°C /
446°F

400°C / 752°F

6000 - 08

72

12

640

1455

6000 - 10

68

6

640

1455

6000 - 15

68

6

640

1455

6000 - 25

42

12

342

755

6000 - 50

42

52

342

755

6000 - 80

42

52

443

960

6000 - 100

42

54

342

755

6000 - 150

42

54

375

852

6000 - 200

42

54

375

852

6000 - 250

42

54

375

852

Resistance (Ohm)

Pt500

Compensation lines

red/brown

brown/orange

6000 - all sizes

540 at 20°C / 68°F

0.1

INFORMATION!
Failure of two or more of the upper circuits may indicate a measuring tube error. There may be
product in the housing. If this is the case, depressurise

depressurise

depressurise

depressurise the process line and immediately remove

the measuring device from the process line.