beautypg.com

Operation – KROHNE MFC 400 EN User Manual

Page 95

background image

OPERATION

6

95

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

6.5.6 Density mode (menu C1.2.2)

There are 4 available operating modes for density that can be set here:

• Process:

The device measures and displays the current operating density of the product.

• Fixed:

The device displays a fixed density value. This value must be entered in menu C1.2.3.

• Referred:

The device calculates the density based on a set reference temperature.

• Standard:

The device calculates the standard density based on a set reference temperature and the
correction factors k

0

...k

2

, so that a standard volume can be output.

The following equation is used for the "Referred" mode:

The following equation is used for the "Referred" mode:

The following equation is used for the "Referred" mode:

The following equation is used for the "Referred" mode:

ρ

r

= ρ

a

+ a (t

a

- t

r

)

ρ

r

= Density at reference temperature

ρ

a

= Current measured operating density at current operating temperature

a = Programmed temperature coefficient/density gradient
t

a

= Current measured operating temperature

t

r

= Reference temperature

The reference temperature must be entered in menu item C1.2.3. The density gradient is set in
C1.2.4.

The following equation is used to calculate the density gradient:

The following equation is used to calculate the density gradient:

The following equation is used to calculate the density gradient:

The following equation is used to calculate the density gradient:

a = (ρ

D

- ρ

C

) / (T

B

- T

A

)

The value for the density gradient is generally positive as an increase in temperature usually
reduces the measured density(exception: water anomaly).

Figure 6-7: Calculating the density gradient

1 Density
2 Temperature

.book Page 95 Monday, August 26, 2013 9:05 AM