Operation, 7 function tests and troubleshooting – KROHNE MFC 400 EN User Manual
Page 113
OPERATION
6
113
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
6.7 Function tests and troubleshooting
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Stores the minimum and maximum value for the temperature occurring during sensor operating
time.
Application problems that may indicate errors in the signal converter:
• An improperly closed shut-off valve during zero calibration results in high calibration values
• Air/gas bubbles result in high energy levels and high calibration values
• Product deposits on the inside of the measuring tube lead to higher/lower density indications
and high calibration values
Common errors (with corresponding symptoms):
• Slightly eroded or corroded measuring tube
- Erroneous density measurement
- High frequency
- Measurement error at low mass flow
• Eroded or corroded measuring tube (product in the housing)
- Measuring tube does not start
- Low resistance to earth with conducting product
• Open circuits of drive and sensor coils or resistance thermometers (RTD)
- Can be measured with ohmmeter
Measuring sensor type
Operating temperature
Minimum
Maximum
OPTIMASS 6000 - Standard
version
-70°C / -94°F
+230°C / +446°F
OPTIMASS 6000 - Cryogenic
version
-200°C / -328°F
+40°C / +104°F
OPTIMASS 6000 - High
temperature version
-50°C / -58°F
+400°C / +752°F
OPTIMASS 6000 - Hastelloy C
-70°C / -94°F
+230°C / +446°F
OPTIMASS 6000 - Duplex
-50°C / -58°F
+230°C / +446°F
.book Page 113 Monday, August 26, 2013 9:05 AM