Technical data – KROHNE MFC 400 EN User Manual
Page 128

8
TECHNICAL DATA
128
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
Passive
U
ext
≤ 32 VDC
-
f
max
in operating menu set to
f
max
≤ 100 Hz:
I ≤ 100 mA
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 0.2 V at I ≤ 10 mA
U
0, max
= 2 V at I ≤ 100 mA
f
max
in operating menu set to
100 Hz < f
max
≤ 10 kHz:
I ≤ 20 mA
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 1.5 V at I ≤ 1 mA
U
0, max
= 2.5 V at I ≤ 10 mA
U
0, max
= 5.0 V at I ≤ 20 mA
NAMUR
-
Passive to EN 60947-5-6
open:
I
nom
= 0.6 mA
closed:
I
nom
= 3.8 mA
Passive to EN 60947-5-6
open:
I
nom
= 0.43 mA
closed:
I
nom
= 4.5 mA
U
i
= 30 V
I
i
= 100 mA
P
i
= 1 W
C
i
=10 nF
L
i
~ 0 mH
Low flow cut off
Low flow cut off
Low flow cut off
Low flow cut off
Function
Switching point and hysteresis separately adjustable for each output, counter and
the display
Switching point
Set in increments of 0.1%.
0…20% (current output, frequency output)
Hysteresis
Set in increments of 0.1%.
0…5% (current output, frequency output)
Time constant
Time constant
Time constant
Time constant
Function
The time constant corresponds to the elapsed time until 67% of the end value has
been reached according to a step function.
Settings
Set in increments of 0.1s.
0…100 s
.book Page 128 Monday, August 26, 2013 9:05 AM