ELECTRICAL CONNECTIONS

4

51

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

4.10.5 Ex i inputs/outputs

Current output active (only current output terminals C/C- have HART

®

capability), Ex i

I/Os

• Observe connection polarity.
• U

int, nom

= 20 VDC

• I ≤ 22 mA
• R

L

≤ 450 Ω

• X designates the connection terminals A or C, depending on the version of the signal

converter.

Current output passive (only current output terminals C/C- have HART

®

capability),

Ex i I/Os

• Any connection polarity.
• U

ext

≤ 32 VDC

• I ≤ 22 mA
• U

0

≥ 4 V

• R

L

≤ (U

ext

- U

0

) / I

max

• X designates the connection terminals A or C, depending on the version of the signal

converter.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

INFORMATION!
For further information on electrical connection refer to Description of the inputs and outputs on
page 38.