Operation – KROHNE MFC 400 EN User Manual

Page 108

background image

6

OPERATION

108

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

S

S Electronics

Electr.Temp.A Out of Spec

Temperature of the signal

converter electronics is out

of range.

Protect signal converter

from process effects and

sunlight.

Electr.Temp.C Out of Spec

Electr.Temp. Out of Spec

S

S Configuration

IO A Overrange

The output value is limited

by a filter.

When limiting the input

value, refer to the actions

listed there. Check the

range setting of the output.

IO B Overrange

IO C Overrange

IO D Overrange

S

S Process

Proc.Temp.Out of Range

The process temperature is

out of range. The

measurement continues but

the accuracy is

undetermined.

Check process conditions.

Flow Out of Range

The flow is out of range. The

actual flow is higher than

the displayed value.

Prod.Density Out of Range

The density is out of range.

The measurement continues

but the accuracy of the

density and flow is

undetermined.

Check process conditions

and density calibration.

Velocity Out of Range

The flow is out of range. The

actual flow is higher or

lower than the value

displayed.

Check process conditions

and pipe diameter setting.

M

M Sensor

M

M Electronics

Backplane Data Faulty

The backplane data record is

faulty.

Check the correct

installation of the signal

converter electronics. After

changing one parameter,

the message should

disappear within one

minute. If not, contact the

manufacturer.

Factory Data Faulty

Factory settings are invalid.

Contact manufacturer.

Backplane Difference

The data on the backplane

differ from the data in the

device.

After changing one

parameter, the message

should disappear within one

minute. If not, contact the

manufacturer.

Mem. Life Exceeded

The number of maximum

write cycles is exceeded for

the memory component.

Exchange the signal

converter electronics.

.book Page 108 Monday, August 26, 2013 9:05 AM