Operation – KROHNE MFC 400 EN User Manual

Page 114

background image

6

OPERATION

114

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

Typical frequency values (at 20°C / 68°F; ±5 Hz)

Problems with zero calibration

1 Stop the flow.
2 Set the totaliser in Fct. C3.y.1 to "Absolute Total".
3 Set the low flow cutoff in Fct. C3.y.3 to zero.
4 Perform automatic zero calibration.
5 Reset totaliser and count for 2 minutes.
6 Compare the added value to the specified zero calibration stability.

Possible causes of poor zero calibration:

• Valves not completely closed, air or gas bubbles or deposits on measuring tube

Size

All materials

Empty

Water

6000 - 08

220

205

6000 - 10

187

172

6000 - 15

212

191

6000 - 25

266

231

6000 - 50

262

230

6000 - 80

234

203

6000 - 100

257

221

6000 - 150

245

208

6000 - 200

218

184

6000 - 250

187

160

INFORMATION!
For best results, perform zero calibration with the medium at operating temperature.

.book Page 114 Monday, August 26, 2013 9:05 AM