beautypg.com

Electrical connections – KROHNE MFC 400 EN User Manual

Page 42

background image

4

ELECTRICAL CONNECTIONS

42

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

Status output / limit switch passive, basic I/Os

• U

ext

≤ 32 VDC

• I ≤ 100 mA
• R

L, max

= 47 kΩ

R

L, min

= (U

ext

- U

0

) / I

max

• open:

I ≤ 0.05 mA at U

ext

= 32 VDC

closed:
U

0, max

= 0.2 V at I ≤ 10 mA

U

0, max

= 2 V at I ≤ 100 mA

• The output is open when the device is de-energized.
• X stands for the connection terminals B, C or D. The functions of the connection terminals

depend on the settings refer to

Function tables on page 68.

Control input passive, basic I/Os

• 8 V ≤ U

ext

≤ 32 VDC

• I

max

= 6.5 mA at U

ext

≤ 24 VDC

I

max

= 8.2 mA at U

ext

≤ 32 VDC

• Switching point for identifying "contact open or closed":

Contact open (off): U

0

≤ 2.5 V with I

nom

= 0.4 mA

Contact closed (on): U

0

≥ 8 V with I

nom

= 2.8 mA

• Can also be set as a status output; for the electrical connection refer to status output

connection diagram.

INFORMATION!

Any connection polarity.

Figure 4-9: Status output / limit switch passive S

p

Figure 4-10: Control input passive C

p

1 Signal

.book Page 42 Monday, August 26, 2013 9:05 AM