OPERATION

6

105

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

Legend

Fixed status signal
Variable status signal

Error
type

Event group

Single event

Description

Actions to eliminate the
event

F

F

F

F

F Sensor

Sensor Containment Dam.

High pressure detected in

housing. Faulty measuring

tube.

Exchange measuring

sensor.

Sensor Error

The measuring sensor

signals are out of range.

Flow measurements are not

possible.

Check connection between

measuring sensor and

signal converter (remote

version) or replace

measuring sensor.

F

F

F

F

F Electronics

System Error

Electronics error in internal

bus communication or due

to a hardware error.

Perform cold start. If

message reappears, contact

manufacturer.

System Error A

System Error C

HW Combination Error

BM Failure

DM Failure

Process Input Failure

IO 1 Failure

IO 2 Failure

Tot 1 Failure

Tot 2 Failure

Tot 3 Failure

IO A Failure

IO B Failure

IO C Failure

F

F

F

F

F Configuration

BM Configuration

Error detected when

starting device. Possible

causes: inadmissible

parameter settings or fault

with electronics component.

Check settings of

appropriate function or load

factory settings. If the error

persists, contact

manufacturer.

DM Configuration

Process Input Config.

Settings for process input

invalid.

Check settings for process

input or load factory

settings.

Density Calib.

Density calibration

parameters invalid.

Perform density calibration.

Check product and function

mode.