OPERATION

6

87

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

6.4.4 Set free units

C6.8 HART

C6.8

HART

C6.8.1

HART

Switch HART

®

communication on/off:

Select:
On (HART

®

activated) possible current range for current output 4…20 mA /

Off (HART

®

not activated) possible current range for current output

0…20 mA

C6.8.2

Address

Set address for HART

®

operation:

Select:

00 (Point-to-Point operation,;current output has normal function;

current = 4…20 mA) / 01...15 (Multi-Drop operation; current output has a

constant setting of 4 mA)

C6.8.3

Message

Set required text:

A…Z ; a…z ; 0…9 ; / - + , . *

C6.8.4

Description

Set required text:

A…Z ; a…z ; 0…9 ; / - + , . *

C6.8.5

HART Long tag

Set the HART Long tags

C6.9 Quick Setup

C6.9

Quick Setup

Activate quick access in the menu "Quick Setup"

Select: Yes (activated) / No (switched off)

C6.9.1

Reset Totaliser 1

Reset Totaliser 1 in menu "Quick Setup"

Select: Yes (activated) / No (switched off)

C6.9.2

Reset Totaliser 2

Reset Totaliser 2 in menu "Quick Setup"

Select: Yes (activated) / No (switched off)

C6.9.3

Reset Totaliser 3

Reset Totaliser 3 in menu "Quick Setup"

Select: Yes (activated) / No (switched off)

C6.9.4

Reset All Totalisers

Reset all Totaliser in menu "Quick Setup"

Select: Yes (activated) / No (switched off)

Free units

Sequences to set texts and factors

Texts

Texts

Texts

Texts

Volume flow, mass flow and density:

3 digits before and after the slash

xxx/xxx (max. 6 characters plus a "/")

Permissible characters:

A…Z; a…z; 0…9; / - + , . *; @ $ % ~ () [] _

Conversion factors

Conversion factors

Conversion factors

Conversion factors

Desired unit

= [unit see above] * conversion factor

Conversion factor

Max. 9 digits

Shift decimal point:

↑ to the left and ↓ to the right

.book Page 87 Monday, August 26, 2013 9:05 AM