Operation – KROHNE MFC 400 EN User Manual
Page 83
OPERATION
6
83
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
No.
Function
Settings / descriptions
C6 Device
C6.1 Device Info
C6.1
Device Info
-
C6.1.1
Tag
Settable characters (max. 8 digits):
A…Z; a…z; 0…9; / - + , .
C6.1.2
V No. Sensor
V number of the measuring sensor, cannot be changed
C6.1.3
Serial No. Sensor
Serial no. of the measuring sensor, cannot be changed
C6.1.4
Sensor Revision
Revision of the measuring sensor, cannot be changed
C6.1.5
V No. Converter
V number of the signal converter, cannot be changed
C6.1.6
Converter Serial No.
Serial no. of the signal converter, cannot be changed
C6.1.7
C Number
CG number, cannot be changed; describes the signal converter version
C6.1.8
Electronics Serial No.
Serial no. of the electronic assembly, cannot be changed
C6.1.9
Electronic Revision
Shows ID no., electronic revision no. and production date;
Contains all hardware and software changes.
C6.2 Display
C6.2
Display
-
C6.2.1
Language
Language selection depends on the device version.
C6.2.2
Contrast
Adjust display contrast for extreme temperatures.
Setting: -9…0…+9
This change takes place immediately, not just when setting mode is exited!
C6.2.3
Default Display
Specification of the default display page that is returned to after a short
delay period.
Select:
None (the current page is always active) / 1st Meas. Page (show this page) /
2nd Meas. Page (show this page) / Status Page (show only status
messages) / Graphic Page (trend display of the 1st measurement)
C6.2.4
Optical Keys
Activate or deactivate the optical keys
Select: On / Off
C6.3 & C6.4 1st Meas. Page & 2nd Meas. Page
C6.3
1st Meas. Page
_ stands for 3 = 1st Meas. Page and 4 = 2nd Meas. Page
C6.4
2nd Meas. Page
C6._.1
Function
Specify number of measured value lines (font size)
Select:
One Line / Two Lines / Three Lines
C6._.2
1st Line Variable
Specify measurement for 1st line
Select:
Volume Flow / Mass Flow / Temperature / Density / Velocity / Sensor
Average / Sensor Deviation / Drive Level / Tube Frequency / Strain 1 /
Strain 2 / 2 Phase Signal
Depending on the settings for the concentration measurement, the following
measurements are possible:
Concentration 1 / Concentration 2 / Conc. mass flow 1 / Conc. mass flow 2 /
Conc. volume flow 1 / Conc. volume flow 2
C6._.3
Range
0…100% of the measurement set in Fct. C5._.2
x.xx…xx.xx _ _ _ (format and unit depend on the "measurement")
.book Page 83 Monday, August 26, 2013 9:05 AM