C1.2.4
Ref. Density Slope
Set the slope for the reference density option
Appears only if density mode "Referred" has been selected in Fct. C1.2.2.
C1.2.5
Standard Density Temp.
Appears only if density mode "Standard" has been selected in Fct. C1.2.2.
Density reference temperature to calculate standard density
C1.2.6
Standard Density k0
Appears only if density mode "Standard" has been selected in Fct. C1.2.2.
Coefficient k
0
to calculate standard density
C1.2.7
Standard Density k1
Appears only if density mode "Standard" has been selected in Fct. C1.2.2.
Coefficient k
1
to calculate standard density
C1.2.8
Standard Density k2
Appears only if density mode "Standard" has been selected in Fct. C1.2.2.
Coefficient k
2
to calculate standard density
C1.2.9
Last Density Cal.
Date of last density calibration
C1.3 Filter
C1.3
Filter
-
C1.3.1
Flow Direction
Define polarity of flow direction.
Forwards (according to the arrow on the measuring sensor) or backwards
(in the opposite direction to the arrow)
C1.3.2
Press. Supp. Time
Set the pressure suppression time; range: 0.0...20.0 s
C1.3.3
Press. Supp. Cutoff
Low flow cutoff settings for the pressure suppression; range: 0.0...10.0%
C1.3.4
Low Flow Cutoff
Set the low flow cutoff; range: 00.0...10.0%
C1.4 System Control
C1.4
System Control
-
C1.4.1
Function
Set the system control.
Select: Inactive (off) / Flow = 0 (flow to zero)
C1.4.2
Sys. Ctrl. Condition
Set the condition for activating the system control.
Selection: density or temperature
C1.4.3
Sys. Ctrl. Max Limit
Defines the upper limit for the condition selected in C1.4.2
C1.4.4
Sys. Ctrl. Min Limit
Defines the lower limit for the condition selected in C1.4.2
C1.5 Diagnosis
C1.5
Diagnosis
-
C1.5.1
Max. Rec. Temp.
Display of maximum recorded sensor temperature
C1.5.2
Min. Rec. Temp.
Display of minimum recorded sensor temperature
C1.5.3
2 Ph. Threshold
Defines the process-dependent sensitivity for 2 phase signal error message.
C1.5.4
Defines the parameter for the respective diagnostic value.
Select:
Off (goes to zero) / Sensor Average (sensor amplitude A+B) / Sensor
Deviation / Drive Level / MT Frequency / Strain 1 / Strain 2 /
2 Phase Signal
C1.5.5
C1.5.6
C1.5.7
Proc: Signal Low
Selection of status signal (Failure, Out of Specification, Function Check,
Maintenance Required and Information) for the group "Proc: Signal Low"
C1.5.8
Proc: Signal Search
Selection of status signal (Failure, Out of Specification, Function Check,
Maintenance Required and Information) for the group "Proc: Signal Search"
C1.5.9
Proc: Current Input
Selection of status signal (Failure, Out of Specification, Function Check,
Maintenance Required and Information) for the group "Proc: Current Input"
C1.5.10
Proc: 2 Phase Flow
Selection of status signal (Failure, Out of Specification, Function Check,
Maintenance Required and Information) for the group "Proc: 2 Phase Flow"
No.
Function
Settings / descriptions