OPERATION

72

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

C1.2.4

Ref. Density Slope

Set the slope for the reference density option

Appears only if density mode "Referred" has been selected in Fct. C1.2.2.

C1.2.5

Standard Density Temp.

Appears only if density mode "Standard" has been selected in Fct. C1.2.2.

Density reference temperature to calculate standard density

C1.2.6

Standard Density k0

Appears only if density mode "Standard" has been selected in Fct. C1.2.2.

Coefficient k

0

to calculate standard density

C1.2.7

Standard Density k1

Appears only if density mode "Standard" has been selected in Fct. C1.2.2.

Coefficient k

1

to calculate standard density

C1.2.8

Standard Density k2

Appears only if density mode "Standard" has been selected in Fct. C1.2.2.

Coefficient k

2

to calculate standard density

C1.2.9

Last Density Cal.

Date of last density calibration

C1.3 Filter

C1.3

Filter

-

C1.3.1

Flow Direction

Define polarity of flow direction.

Forwards (according to the arrow on the measuring sensor) or backwards

(in the opposite direction to the arrow)

C1.3.2

Press. Supp. Time

Set the pressure suppression time; range: 0.0...20.0 s

C1.3.3

Press. Supp. Cutoff

Low flow cutoff settings for the pressure suppression; range: 0.0...10.0%

C1.3.4

Low Flow Cutoff

Set the low flow cutoff; range: 00.0...10.0%

C1.4 System Control

C1.4

System Control

-

C1.4.1

Function

Set the system control.

Select: Inactive (off) / Flow = 0 (flow to zero)

C1.4.2

Sys. Ctrl. Condition

Set the condition for activating the system control.

Selection: density or temperature

C1.4.3

Sys. Ctrl. Max Limit

Defines the upper limit for the condition selected in C1.4.2

C1.4.4

Sys. Ctrl. Min Limit

Defines the lower limit for the condition selected in C1.4.2

C1.5 Diagnosis

C1.5

Diagnosis

-

C1.5.1

Max. Rec. Temp.

Display of maximum recorded sensor temperature

C1.5.2

Min. Rec. Temp.

Display of minimum recorded sensor temperature

C1.5.3

2 Ph. Threshold

Defines the process-dependent sensitivity for 2 phase signal error message.

C1.5.4

Defines the parameter for the respective diagnostic value.

Select:

Off (goes to zero) / Sensor Average (sensor amplitude A+B) / Sensor

Deviation / Drive Level / MT Frequency / Strain 1 / Strain 2 /

2 Phase Signal

C1.5.5

C1.5.6

C1.5.7

Proc: Signal Low

Selection of status signal (Failure, Out of Specification, Function Check,

Maintenance Required and Information) for the group "Proc: Signal Low"

C1.5.8

Proc: Signal Search

Selection of status signal (Failure, Out of Specification, Function Check,

Maintenance Required and Information) for the group "Proc: Signal Search"

C1.5.9

Proc: Current Input

Selection of status signal (Failure, Out of Specification, Function Check,

Maintenance Required and Information) for the group "Proc: Current Input"

C1.5.10

Proc: 2 Phase Flow

Selection of status signal (Failure, Out of Specification, Function Check,

Maintenance Required and Information) for the group "Proc: 2 Phase Flow"

No.

Function

Settings / descriptions