6.4.3 Menu C, Setup

B3 Information

B3

Information

B3.1

Status Log

Log with date and time of status messages that occurred

B3.2

Status Details

Details about the status messages displayed

B3.3

C Number

CG number, cannot be changed (input/output version)

B3.4

Sensor Electronics

Displays the version of the measuring sensor electronics

B3.5

"Bus interface"

Only appears with Profibus, Modbus and FF

B3.6

Electronic Revision

For presentation of LC display see Fct. B3.3 and B3.4

B3.7

Sensor Revision

Revision of measuring sensor

B3.8

Change Log

The check sum across all settings is displayed directly below the menu item.

Shows the changes including date, time and check sum for all settings and
parameters in the converter, regardless of the interface (display, HART

®

,

PROFIBUS, ...) used.

No.

Function

Settings / descriptions

C1 Process Input

C1.1 Zero and Offset

C1.1

Zero and Offset

-

C1.1.1

Zero Calibration

Start zero calibration

Query: calibrate zero?

Setting:

Automatic / Factory Settings /

Manual (display last value; set new value, range: -10…+10%) /

Break (return with ^ key)

C1.1.2

Zero Add. Offset

Direct setting of zero offset

C1.1.3

Pipe Diameter

Set the pipe diameter in mm to calculate flow velocity

C1.1.4

Flow Correction

Defines additional correction for mass flow;

Range: -100...+100%

C1.1.5

Zero Calibration Log

Log of last zero calibration including zero point, temperature, date and time

C1.2 Density

C1.2.1

Density Calib.

Start density calibration

For detailed information refer to

Density calibration (menu C1.2.1) on page

90.

C1.2.2

Density Mode

Selection of density mode:

Process (return with ^ key /

Fixed (a fixed value is used for the density (e.g. standard density)) /

Referred (calculates process density based on a reference temperature) /

Standard (calculates standard density based on a reference temperature

and the correction factors)

C1.2.3

Fixed Density Value

Set the fixed value (e.g. standard density) for the density.

Appears only if density mode "Fixed" has been selected in Fct. C1.2.2.

C1.2.3

Density Ref. Temp.

Set the reference temperature for the reference density option

Appears only if density mode "Referred" has been selected in Fct. C1.2.2.