OPERATION

6

109

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

M

M

Configuration

Backup 1 Data Faulty

Error when checking the

data record for Backup 1.

Use "Setup > Device >

Special Functions > Save

Settings" to save the data

record. If the message

continues to appear, contact

the manufacturer.

Backup 2 Data Faulty

Error when checking the

data record for Backup 2.

M

M Process

F

F

F

F

F Proc: Current

Input

IO A Error Current

Error current is detected at

current input A.

Check device connected to

current input A. This devices

submits an error current.

IO B Error Current

Error current is detected at

current input B.

Check device connected to

current input B. This devices

submits an error current.

S

S Electr: IO

Connection

IO A Connection

Current output A cannot

provide the necessary

current. The current

provided is too low. The

current at input A is below

0.5 mA or above 23 mA.

Check connection at A.

Measure resistance of

current loop at A.

Check current at A.

IO A Connection

Open circuit or short circuit

at IO A.

IO B Connection

Current output B cannot

provide the necessary

current. The current

provided is too low. The

current at input B is below

0.5 mA or above 23 mA.

Check connection at B.

Measure resistance of

current loop at B.

Check current at B.

IO B Connection

Open circuit or short circuit

at IO B.

IO C Open Connector

Current output C cannot

provide the necessary

current.

The current provided is too

low.

Check connection at C.

Measure resistance of

current loop at C.

F

F

F

F

F Proc: Signal

Search

Sensor signal search

The measuring sensor

signals are too small for an

accurate flow measurement.

The cause is typically too

much damping, e.g. due to

gas bubbles or the viscosity

is too high.

Check process conditions. In

the case of a measuring

sensor in remote version,

check the electrical

connection between the field

housing and the measuring

sensor.

S

S Proc: 2 Phase

Flow

2 Phase Flow Detected

2 phase flow has been

detected. The accuracy of

the flow and density

measurement is

undetermined.

Check product for gas

bubbles.

.book Page 109 Monday, August 26, 2013 9:05 AM