Operation – KROHNE MFC 400 EN User Manual
Page 109
OPERATION
6
109
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
M
M
Configuration
Backup 1 Data Faulty
Error when checking the
data record for Backup 1.
Use "Setup > Device >
Special Functions > Save
Settings" to save the data
record. If the message
continues to appear, contact
the manufacturer.
Backup 2 Data Faulty
Error when checking the
data record for Backup 2.
M
M Process
F
F
F
F
F Proc: Current
Input
IO A Error Current
Error current is detected at
current input A.
Check device connected to
current input A. This devices
submits an error current.
IO B Error Current
Error current is detected at
current input B.
Check device connected to
current input B. This devices
submits an error current.
S
S Electr: IO
Connection
IO A Connection
Current output A cannot
provide the necessary
current. The current
provided is too low. The
current at input A is below
0.5 mA or above 23 mA.
Check connection at A.
Measure resistance of
current loop at A.
Check current at A.
IO A Connection
Open circuit or short circuit
at IO A.
IO B Connection
Current output B cannot
provide the necessary
current. The current
provided is too low. The
current at input B is below
0.5 mA or above 23 mA.
Check connection at B.
Measure resistance of
current loop at B.
Check current at B.
IO B Connection
Open circuit or short circuit
at IO B.
IO C Open Connector
Current output C cannot
provide the necessary
current.
The current provided is too
low.
Check connection at C.
Measure resistance of
current loop at C.
F
F
F
F
F Proc: Signal
Search
Sensor signal search
The measuring sensor
signals are too small for an
accurate flow measurement.
The cause is typically too
much damping, e.g. due to
gas bubbles or the viscosity
is too high.
Check process conditions. In
the case of a measuring
sensor in remote version,
check the electrical
connection between the field
housing and the measuring
sensor.
S
S Proc: 2 Phase
Flow
2 Phase Flow Detected
2 phase flow has been
detected. The accuracy of
the flow and density
measurement is
undetermined.
Check product for gas
bubbles.
.book Page 109 Monday, August 26, 2013 9:05 AM