Operation – KROHNE MFC 400 EN User Manual
Page 111
OPERATION
6
111
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
I
I Config: No
Meas.Value
IO A No Meas. Values
The measurement value for
the IO terminals A is not
active.
Select another
measurement for IO A in
menu C3.2 or activate the
measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
IO B No Meas. Values
The measurement value for
the IO terminals B is not
active.
Select another
measurement for IO B in
menu C3.3 or activate the
measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
IO C No Meas. Values
The measurement value for
the IO terminals C is not
active.
Select another
measurement for IO C in
menu C3.4 or activate the
measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
IO D No Meas. Values
The measurement value for
the IO terminals D is not
active.
Select another
measurement for IO D in
menu C3.5 or activate the
measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
Tot 1 No Meas. Values
The measurement value for
the totaliser 1 is not active.
Select another
measurement for totaliser 1
in menu C4.1 or activate the
measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.
Tot 2 No Meas. Values
The measurement value for
the totaliser 2 is not active.
Select another
measurement for totaliser 2
in menu C4.2 or activate the
measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.
Tot 3 No Meas. Values
The measurement value for
the totaliser 3 is not active.
Select another
measurement for totaliser 3
in menu C4.3 or activate the
measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.
Disp. 1 No Meas. Values
Measurement value at 1st
measuring page is not
active.
Select another
measurement for 1st
measuring page or activate
the measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.
Disp. 2 No Meas. Values
Measurement value at 2nd
measuring page is not
active.
Select another
measurement for 2nd
measuring page or activate
the measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.