6

OPERATION

110

MFC 400

www.krohne.com

08/2013 - 4002075802 - MA MFC 400 R02 en

S

S Proc: Signal

Low

Sensor signal low

The sensor signals do not

reach the expected

amplitude. This can

influence the measuring

accuracy.

Check the product for

viscosity and gas bubbles.

S

S Config:

Totaliser

Tot 1 FB4 Overflow

Totaliser 1 FB4 has

overflowed and started

again at zero.

Check totaliser format.

Tot 2 FB5 Overflow

Totaliser 2 FB5 has

overflowed and started

again at zero.

Tot 3 FB6 Overflow

Totaliser 3 FB6 has

overflowed and started

again at zero.

Tot 1 Overflow

Totaliser 1 has overflowed

and started again at zero.

Tot 2 Overflow

Totaliser 2 has overflowed

and started again at zero.

Tot 3 Overflow

Totaliser 3 has overflowed

and started again at zero.

I

I Proc: System

Control

System Control Active

System control is active.

I

I Electr: Power

Failure

Tot 1 Power Failure

Totaliser 1 invalid due to

power failure.

Check the value of the

totaliser.

Tot 2 Power Failure

Totaliser 2 invalid due to

power failure.

Tot 3 Power Failure

Totaliser 3 invalid due to

power failure.

Tot 1 FB4 Power Failure

Totaliser 1 FB4 invalid due

to power failure.

Tot 2 FB5 Power Failure

Totaliser 2 FB5 invalid due

to power failure.

Tot 3 FB6 Power Failure

Totaliser 3 FB6 invalid due

to power failure.

Power Failure Detected

Power failure has occurred.

The totaliser state may be

invalid.

.book Page 110 Monday, August 26, 2013 9:05 AM