Technical data – KROHNE MFC 400 EN User Manual
Page 126
8
TECHNICAL DATA
126
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
Inputs and outputs
General
All outputs are electrically isolated from each other and from all other circuits.
All operating data and output values can be adjusted.
Description of abbreviations
U
ext
= external voltage; R
L
= load + resistance;
U
0
= terminal voltage; I
nom
= nominal current
Safety limit values (Ex i):
U
i
= max. input voltage; I
i
= max. input current; P
i
= max. input power rating;
C
i
= max. input capacity; L
i
= max. input inductivity
Current output
Current output
Current output
Current output
Output data
Volume flow, mass flow, temperature, density, flow velocity, diagnostic value, 2-
phase signal
Concentration and concentration flow are also possible with available concentration
measurement (optional).
Temperature coefficient
Typically ±30 ppm/K
Settings
Without HART
Without HART
Without HART
Without HART
®
Q = 0%: 0…20 mA; Q = 100%: 10…20 mA
Error identification: selectable 3…22 mA
With HART
With HART
With HART
With HART
®
Q = 0%: 4…20 mA; Q = 100%: 10…20 mA
Error identification: selectable 3…22 mA
Operating data
Basic I/Os
Basic I/Os
Basic I/Os
Basic I/Os
Modular I/Os
Modular I/Os
Modular I/Os
Modular I/Os
Ex i
Ex i
Ex i
Ex i
Active
U
int, nom
= 24 VDC
I ≤ 22 mA
R
L
≤ 1 kΩ
U
int, nom
= 20 VDC
I ≤ 22 mA
R
L
≤ 450 Ω
U
0
= 21 V
I
0
= 90 mA
P
0
= 0.5 W
C
0
= 90 nF / L
0
= 2 mH
C
0
= 110 nF / L
0
= 0.5 mH
Passive
U
ext
≤ 32 VDC
I ≤ 22 mA
U
0
≥ 1.8 V
R
L
≤ (U
ext
- U
0
) / I
max
U
ext
≤ 32 VDC
I ≤ 22 mA
U
0
≥ 4 V
R
L
≤ (U
ext
- U
0
) / I
max
U
i
= 30 V
I
i
= 100 mA
P
i
= 1 W
C
i
= 10 nF
L
i
~ 0 mH
.book Page 126 Monday, August 26, 2013 9:05 AM