21 care & maintenance – In-Situ TROLL 9500 Operators Manual User Manual
Page 146
141
Multi-Parameter
Water Quality TROLL
®
TROLL 9500 Operator’s Manual
0095110 rev. 007 01/09
21 CARE & MAINTENANCE
REPLACING BATTERIES
The MP TROLL 9500 uses—
s
optical dissolved oxygen sensor, and with a turbidity wiper
Use only Saft LSH-20 3.6V lithium D cells. Use of any
other lithium battery will void the product warranty.
Battery voltage and approximate percentage remaining is displayed
in the software interface when the instrument is connected to a PC.
Note: Due to the voltage supplied by two lithium D cells, the software
may report that the TROLL 9500 is operating on external power.
To replace batteries:
1. Unscrew and remove the white battery compart-
ment cover. If the cable is attached, slide the
battery compartment cover up onto the cable.
2. Press down slightly on the top battery to remove it,
or knock it out gently into your hand.
3. Tip the unit to slide the bottom battery out.
4. Insert the new batteries according to the dia-
gram on the inside of the battery compartment
(positive up for both).
5. Slide the white cover back down over the battery compartment and
hand-tighten to thread it to the instrument body.
Screw the cover down firmly to compress the o-rings and
create a waterproof seal. When properly assembled, the
o-rings will not be vsible.
6. At your next software connection in Win-Situ 4 or Pocket-Situ 4,
edit the device to update the battery information. (See “Editing the
Device Properties” in Section 4.)
O-RING SEALS
LUBRICATION
The Viton
®
o-rings used in the MP TROLL 9500 and other submersible
In-Situ instruments are crucial to insure the integrity of the water-
tight seal. We recommend that you inspect them each time they are
stressed (insertion/removal of sensors, attachment/removal of the
restrictor, battery replacement, etc.) for any indication of dirt, cracks,
tears, splitting, shredding, desiccation, and other damage. If the o-
rings are in good condition, apply silicone lubricant before re-assem-
bling the instrument. Remove excess lubricant with a lint-free tissue.
If the o-rings are damaged, they should be replaced, as described
below.
When lubricating the sensor o-rings, take special care to keep grease
away from the area around the connector at the bottom of the sen-
sors. Should lubricant get into this area, remove it with a clean cotton
swab.
Insert batteries negative side first,
positive side up