CONFIGURATION

85

Acquisition

Acquisition process samples the analog input signals and converts
them into digital format for internal processing.

Select acquisition mode

Panel operation

1. Press the Acquire key.

Acquire

2. Select the acquisition mode

from F1 (Normal) ~ F3
(Average). The acquisition
icon on the top right corner
of the display changes
accordingly.

Normal

F 1

Peak

Detect

F 2

Average

2

F 3

Range Normal

All of the acquired data is used to

draw the waveform.

Peak
Detect

Only the minimum and maximum

value pairs for each acquisition
interval (bucket) are used. This
mode is useful for catching
abnormal glitches in the signal.

Average

Multiple acquired data are
averaged. This mode is useful for
drawing a noise-free waveform.
To select the average number,
press F3 repeatedly.
Average number: 2, 4, 8, 16, 32, 64,
128, 256