beautypg.com

Program, Overview – GW Instek GDS-2000 series Uer manual User Manual

Page 78

background image

GDS-2000 Series User Manual

78

Program

Overview

Background

Program function measures input signals using
cursors or automatic measurement functions, in
user-defined sequence, duration, loop count, and
panel settings. This feature is useful for automated
and repetitive measurement, such as in assembly
line or quality inspection test.

Parameter Program

set

1 set

Program step

Maximum 20 steps

Measurement
item

Cursor or Automatic
measurement

Time (duration)
per step

1 ~ 99 seconds, or user activation

Program loop

1 ~ 99 loops, the first and last step
settable

Programming
step

1. Show the target waveform on the display and

decide the type of measurement that needs to
be done: Horizontal/Vertical Cursor or
Automatic measurement.

2. Setup the other panel configurations: trigger,

acquisition, horizontal/vertical scale, etc. Save
the settings to the internal memory. See
page129 for details.

3. Edit the program (page79) using the internally

stored panel setup.

4. Run the program (page81).