Stop

The test stops when the NoGo
condition is met. The buzzer
does not sound.

Stop+

The test stops and the buzzer
sounds when the NoGo
condition is met.

Continue

The test continues even when
the NoGo condition is met. The
buzzer does not sound.

Continue+

The test continues even when
the NoGo condition is met. The
buzzer also sounds.

Note

If the sound is turned Off in the buzzer setting
(page70), the sound is not produced even when

selecting Stop/Continue+ .

Edit: Template (boundary)

Background

The NoGo template sets the upper and lower
amplitude boundary. Two methods are available:
Min/Max and Auto.

Min/Max

Selects the upper boundary (Max) and
lower boundary (Min) as separate
waveforms, from the internal memory.
Advantage: The template shape and the
distance (allowance) between the source
signal are fully customizable.
Disadvantage: The waveforms
(templates) have to be stored internally
prior to this selection.