beautypg.com

GW Instek GDS-2000 series Uer manual User Manual

Page 121

background image

SAVE/RECALL

121

Contents:
waveform data

The waveform data can be used for detailed
analysis. It consists of horizontal and vertical
position of the waveform for the entire memory
length.

One division includes
25 points of horizontal
and vertical data. The
vertical point starts
from the center line.
The horizontal point
starts from the leftmost
waveform.

0

25

–25

25

50

75

The time length or voltage level which each data
point represents differs according to the vertical
and horizontal scale. For example:
Vertical scale: 10mV/div ( 4mV per point)
Horizontal scale: 100us/div (4us per point)

Contents: other
data

The following information is also included in the
waveform file.

Memory length

source channel

vertical offset

vertical scale

coupling mode

waveform last dot
address

date and time

trigger level

vertical position

time base

probe attenuation

horizontal view

horizontal scale

sampling period

sampling mode