MEASUREMENT

73

Auto

Creates the upper and lower boundary
together from an input signal, not from
internally stored waveform.
Advantage: No need to store the
waveforms prior to this selection.
Disadvantage: The template shape is
proportional to the source signal. The
distance (allowance) between the source
signal and upper/lower template are
always symmetrical.

Min/Max setting

1. Make sure the source signal, on which the

templates are based, appears on the display.

2. Press the Utility key.

Utility

3. Press F5 (More).

More

F 5

4. Press F3 (Go-NoGo Menu).

Go-NoGo

Menu

F 3

5. Press F1 (Template Edit).

Template

Edit

F 1

6. Press F1 (Template)

repeatedly to select the
upper (Max) or lower (Min)
boundary template.

Template

Max

F 1

7. Press F2 (Source). Use the

Variable knob to select the
template from internally
stored waveform. For
waveform store procedure,
see page130.

Source

RefA

F 2