beautypg.com

Recall reference waveform on the display – GW Instek GDS-2000 series Uer manual User Manual

Page 139

background image

SAVE/RECALL

139

Recall reference waveform on the display

Panel operation

1. The reference waveform must be stored in

advance. See page for waveform store details.

2. Press the Save/Recall key.

Save/Recall

3. Press F2 (Display Refs). The

reference waveform display
menu appears.

Display

Refs.

F 2

4. Select the reference

waveform from F1 (Ref A)
to F4 (Ref D) and press it.
The waveform appears on
the display and the period
and amplitude of the
waveform appears in the
menu.

Ref.A Off

F 2

Ref.A On

F 2

1V
2.5ms

5. To clear the waveform from

the display, press F1 ~ F4
key again.

Ref.A Off

F 2