GW Instek GDS-2000 series Uer manual User Manual

Page 160

background image

GDS-2000 Series User Manual

160

3. Press F5 (More) twice.

More

F 5

x2

4. Press F1 (ProbeComp

Menu).

ProbeComp

Menu

F 1

5. Press F1 (Wavetype)

repeatedly to select the
standard square wave.

6. Press the Auto Set key. The

compensation signal
appears on the display.

Auto Set

7. Press the Display key, then

F1 (Type) twice to select the
vector waveform.

Display

Type

Vectors

F 1

8. Turn the adjustment point on the probe until

the signal edge becomes sharp.

Under

Compensation

Normal

Over

Compensation